Safety and biocompatibility of a novel self-expanding nitinol carotid stent with hybrid cell design in a porcine model of neointimal hyperplasia

Adam Janas, Krzysztof Milewski, Piotr P. Buszman, Przemysław Nowakowski, Michał Jelonek, Bartłomiej Orlik, Agata Krauze, Stefan Samborski, Diane Beaudry, Guy Lecelrc, Marek Król, Jean−Martin Lapointe, Wojciech Wojakowski, Anna Turek, Radosław S. Kiesz, Paweł E. Buszman
DOI: 10.5603/KP.a2014.0219
·
Kardiol Pol 2015;73(4):240-245.

open access

Vol 73, No 4 (2015)
Original articles - new methods
Published online: 2015-04-14
Submitted: 2014-02-23
Accepted: 2014-09-11

Abstract

Background: Stent design may influence the outcomes, suggesting that adverse event rates vary according to free cell area and cell design. Open cell design technology of self-expandable stents, dedicated for carotid revascularisation has better deliverability, although closed cell technology is expected to cause fewer thromboembolic events.

Aim: To evaluate the feasibility and vascular response of novel, hybrid cell, self-expandable nitinol stents (MER®, Balton, Poland) implanted into porcine carotid arteries. Hybrid cell design combines open and closed cell technology.

Methods: All tested stents were implanted with 10% overstretch into 10 carotid segments of Polish domestic pigs. Control angiography was obtained immediately before and after vascular interventions as well as 28 days after the procedure. Thereafter, animals were sacrificed, and the treated segments were harvested and evaluated in the independent histopathology laboratory.

Results: All stents were easily introduced and implanted, showing good angiographic acute outcome. At 28 days, in the angiography, all vessels were patent with no signs of thrombi or excessive neointimal formation, with the late lumen loss of –0.11 ± 0.3 mm and percentage diameter stenosis 10.18 ± 8.1%. There was a 10% increase in the vessel reference diameter when compared to baseline (4.57 ± 0.5 vs. 4.96 ± 0.3 mm, p < 0.01). In the histopathology, mean area stenosis was 17.4% and mean intimal thickness was 0.20 mm. At histopathology, the mean injury, inflammation, and fibrin scores were low. Endothelialisation was complete in all stents, and neointimal tissue appeared moderately mature as shown by the moderate mean neointimal smooth muscle score. Nonetheless, histopathology shows one stent affected by peri-strut granulomas and one stent affected by marked mineralisation.

Conclusions: The novel Polish self-expandable nitinol carotid stent with hybrid cell technology shows optimal biocompatibility and a vascular healing profile, and therefore may be introduced for first-in-man application.

Abstract

Background: Stent design may influence the outcomes, suggesting that adverse event rates vary according to free cell area and cell design. Open cell design technology of self-expandable stents, dedicated for carotid revascularisation has better deliverability, although closed cell technology is expected to cause fewer thromboembolic events.

Aim: To evaluate the feasibility and vascular response of novel, hybrid cell, self-expandable nitinol stents (MER®, Balton, Poland) implanted into porcine carotid arteries. Hybrid cell design combines open and closed cell technology.

Methods: All tested stents were implanted with 10% overstretch into 10 carotid segments of Polish domestic pigs. Control angiography was obtained immediately before and after vascular interventions as well as 28 days after the procedure. Thereafter, animals were sacrificed, and the treated segments were harvested and evaluated in the independent histopathology laboratory.

Results: All stents were easily introduced and implanted, showing good angiographic acute outcome. At 28 days, in the angiography, all vessels were patent with no signs of thrombi or excessive neointimal formation, with the late lumen loss of –0.11 ± 0.3 mm and percentage diameter stenosis 10.18 ± 8.1%. There was a 10% increase in the vessel reference diameter when compared to baseline (4.57 ± 0.5 vs. 4.96 ± 0.3 mm, p < 0.01). In the histopathology, mean area stenosis was 17.4% and mean intimal thickness was 0.20 mm. At histopathology, the mean injury, inflammation, and fibrin scores were low. Endothelialisation was complete in all stents, and neointimal tissue appeared moderately mature as shown by the moderate mean neointimal smooth muscle score. Nonetheless, histopathology shows one stent affected by peri-strut granulomas and one stent affected by marked mineralisation.

Conclusions: The novel Polish self-expandable nitinol carotid stent with hybrid cell technology shows optimal biocompatibility and a vascular healing profile, and therefore may be introduced for first-in-man application.

Get Citation

Keywords

carotid stenting, new technologies, pre-clinical study

About this article
Title

Safety and biocompatibility of a novel self-expanding nitinol carotid stent with hybrid cell design in a porcine model of neointimal hyperplasia

Journal

Kardiologia Polska (Polish Heart Journal)

Issue

Vol 73, No 4 (2015)

Pages

240-245

Published online

2015-04-14

DOI

10.5603/KP.a2014.0219

Bibliographic record

Kardiol Pol 2015;73(4):240-245.

Keywords

carotid stenting
new technologies
pre-clinical study

Authors

Adam Janas
Krzysztof Milewski
Piotr P. Buszman
Przemysław Nowakowski
Michał Jelonek
Bartłomiej Orlik
Agata Krauze
Stefan Samborski
Diane Beaudry
Guy Lecelrc
Marek Król
Jean−Martin Lapointe
Wojciech Wojakowski
Anna Turek
Radosław S. Kiesz
Paweł E. Buszman

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., Świętokrzyska 73 street, 80–180 Gdańsk, Poland

tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail: viamedica@viamedica.pl