Non-functional lead as a cause of detection disturbances in implantable cardioverter-defibrillator

Maciej Kempa, Szymon Budrejko, Mariusz Nowakowski, Tomasz Królak, Ludmiła Daniłowicz−Szymanowicz, Grzegorz Raczak
Kardiol Pol 2012;70(10):1073-1075.

open access

Vol 70, No 10 (2012)
Invasive electrophysiology
Published online: 2012-10-19
Submitted: 2012-12-28

Abstract

The article presents the case of a patient treated with implantable cardioverter-defibrillator, in whom an additional lead (integrated bipolar) was implanted because of the damage to the ventricular lead. Interactions between the electrodes led to detection disturbances. The patient required the removal of the non-functional lead.

Abstract

The article presents the case of a patient treated with implantable cardioverter-defibrillator, in whom an additional lead (integrated bipolar) was implanted because of the damage to the ventricular lead. Interactions between the electrodes led to detection disturbances. The patient required the removal of the non-functional lead.
Get Citation

Keywords

implantable cardioverter-defibrillator; inappropriate therapy

About this article
Title

Non-functional lead as a cause of detection disturbances in implantable cardioverter-defibrillator

Journal

Kardiologia Polska (Polish Heart Journal)

Issue

Vol 70, No 10 (2012)

Pages

1073-1075

Published online

2012-10-19

Bibliographic record

Kardiol Pol 2012;70(10):1073-1075.

Keywords

implantable cardioverter-defibrillator
inappropriate therapy

Authors

Maciej Kempa
Szymon Budrejko
Mariusz Nowakowski
Tomasz Królak
Ludmiła Daniłowicz−Szymanowicz
Grzegorz Raczak

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., Świętokrzyska 73 street, 80–180 Gdańsk, Poland

tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail: viamedica@viamedica.pl