Effects of shock polarity reversal on defibrillation threshold in an implantable cardioverter-defibrillator

Agnieszka Zienciuk, Andrzej Lubiński, Tomasz Królak, Ewa Lewicka-Nowak, Maciej Kempa, Anna Pazdyga, Grzegorz Raczak
Kardiol Pol 2007;65(5):495-500.

open access

Vol 65, No 5 (2007)
Other
Published online: 2007-06-01
Submitted: 2012-12-28

Abstract

Background: An increased defibrillation threshold (DFT) may limit the efficacy of an implantable cardioverter-defibrillator (ICD) in termination of life-threatening ventricular arrhythmias. A search for methods of decreasing DFT has been ongoing since the introduction of ICD into clinical practice. Aim: To assess the effects of various shock polarities on DFT. Methods: The study group consisted of 19 patients (8 females and 11 males, mean age 52±17 years) who received devices (Biotronik, Germany) with a single-coil defibrillation lead. In all patients the value of DFT was assessed using a normal shock polarity as well as using a reversed polarity shock, starting from the energy lower than that measured during normal DFT testing. The impedance of the defibrillation system using two different polarities was also measured. The effects of demographic and clinical parameters on defibrillation parameters were also examined. Results: When using normal shock polarity, the mean DFT value was 12±5 J (range 3.1-20 J) and impedance was 64±12 W. When shock polarity was reversed, the mean DFT value was 9.2±5.0 J (range 2-20 J) and impedance was 67±11 W. In 11 (58%) patients the polarity change caused a marked (by 37%) decrease in the mean DFT value – from 11.5±5.1 J to 7.2±3.8 J. In 5 patients DFT reduction was ł5 J. There was no relationship between demographic or clinical parameters and defibrillation efficacy using the two tested shock polarities. Conclusions: The reversal of shock polarity reduces DFT in more than half of patients. In patients with a high DFT the use of reversed polarity of defibrillating impulse may reduce DFT, which widens the safety margin and makes implantation of additional leads unnecessary. Because clinical parameters have no value in predicting the effects of polarity changes on DFT, the efficacy of reversed polarity shock has to be assessed individually in each patient.

Abstract

Background: An increased defibrillation threshold (DFT) may limit the efficacy of an implantable cardioverter-defibrillator (ICD) in termination of life-threatening ventricular arrhythmias. A search for methods of decreasing DFT has been ongoing since the introduction of ICD into clinical practice. Aim: To assess the effects of various shock polarities on DFT. Methods: The study group consisted of 19 patients (8 females and 11 males, mean age 52±17 years) who received devices (Biotronik, Germany) with a single-coil defibrillation lead. In all patients the value of DFT was assessed using a normal shock polarity as well as using a reversed polarity shock, starting from the energy lower than that measured during normal DFT testing. The impedance of the defibrillation system using two different polarities was also measured. The effects of demographic and clinical parameters on defibrillation parameters were also examined. Results: When using normal shock polarity, the mean DFT value was 12±5 J (range 3.1-20 J) and impedance was 64±12 W. When shock polarity was reversed, the mean DFT value was 9.2±5.0 J (range 2-20 J) and impedance was 67±11 W. In 11 (58%) patients the polarity change caused a marked (by 37%) decrease in the mean DFT value – from 11.5±5.1 J to 7.2±3.8 J. In 5 patients DFT reduction was ł5 J. There was no relationship between demographic or clinical parameters and defibrillation efficacy using the two tested shock polarities. Conclusions: The reversal of shock polarity reduces DFT in more than half of patients. In patients with a high DFT the use of reversed polarity of defibrillating impulse may reduce DFT, which widens the safety margin and makes implantation of additional leads unnecessary. Because clinical parameters have no value in predicting the effects of polarity changes on DFT, the efficacy of reversed polarity shock has to be assessed individually in each patient.
Get Citation

Keywords

ICD; defibrillation threshold; shock polarity

About this article
Title

Original article
Effects of shock polarity reversal on defibrillation threshold in an implantable cardioverter-defibrillator

Journal

Kardiologia Polska (Polish Heart Journal)

Issue

Vol 65, No 5 (2007)

Pages

495-500

Published online

2007-06-01

Bibliographic record

Kardiol Pol 2007;65(5):495-500.

Keywords

ICD
defibrillation threshold
shock polarity

Authors

Agnieszka Zienciuk
Andrzej Lubiński
Tomasz Królak
Ewa Lewicka-Nowak
Maciej Kempa
Anna Pazdyga
Grzegorz Raczak

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., Świętokrzyska 73 street, 80–180 Gdańsk, Poland

tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail: viamedica@viamedica.pl