Technical and practical aspects of remote monitoring of implantable cardioverter-defibrillator patients in Poland – preliminary results

Andrzej Przybylski, Joanna Zakrzewska-Koperska, Aleksander Maciąg, Paweł Derejko, Michał Orczykowski, Łukasz Szumowski, Franciszek Walczak
Kardiol Pol 2009;67(5):505-511.

open access

Vol 67, No 5 (2009)
Other
Published online: 2009-05-20
Submitted: 2012-12-28

Abstract


Background: The aim of remote monitoring of implantable cardioverter-defibrillators (ICD) is to increase the patient’s safety by early detection of technical or medical malfunctions and decrease the number of follow-up visits.
Aim: To evaluate the feasibility and reliability of internet-based home monitoring of ICD recipients in Poland.
Methods: Twenty-seven patients with ICD with remote monitoring options were evaluated; 20 (74%) patients had a single chamber ICD, 6 (22%) patients had a dual chamber ICD and one had an ICD with a resynchronisation therapy option. Medical and technical events reported by the remote monitoring system as well as interruptions in monitoring longer than 14 days were analysed.
Results: The patients were followed for 12.7 ± 10.5 months. Two of them died because of heart failure (6 and 13 months after ICD implantation, respectively). The remote monitoring system reported medical events in 13 (48%) patients. In total, we received 32 event reports (from 1 to 19 per patient, mean 2.6) which were generated due to the detection of ventricular tachycardia (VT) (17 events in 9 patients), ventricular fibrillation (VF) (9 episodes in 6 patients), ineffective defibrillation with the maximal energy (5 reports in 3 patients) and supraventricular tachycardia in the VT detection window (1). Two patients had more than 3 VT/VF episodes during 24 h. There were no reports on technical abnormalities of the ICD system. Interruptions in home monitoring longer than 14 days occurred in 5 (18.5%) patients and lasted 2 to 14 weeks (mean 2.8 ± 7.1). The longest break was caused by the patient’s stay abroad. The remaining interruptions were caused by: journeys (5 episodes), hospitalisations (4), and a temporary stay in a place without sufficient GSM coverage (3). During the follow-up period there were no interruptions in monitoring caused by transmitter or ICD failure. All data received by the home monitoring system were confirmed during the follow-up visits.
Conclusion: Remote monitoring of ICD recipients in Poland does not present technical difficulties and enables early detection of serious events in ICD patients.

Abstract


Background: The aim of remote monitoring of implantable cardioverter-defibrillators (ICD) is to increase the patient’s safety by early detection of technical or medical malfunctions and decrease the number of follow-up visits.
Aim: To evaluate the feasibility and reliability of internet-based home monitoring of ICD recipients in Poland.
Methods: Twenty-seven patients with ICD with remote monitoring options were evaluated; 20 (74%) patients had a single chamber ICD, 6 (22%) patients had a dual chamber ICD and one had an ICD with a resynchronisation therapy option. Medical and technical events reported by the remote monitoring system as well as interruptions in monitoring longer than 14 days were analysed.
Results: The patients were followed for 12.7 ± 10.5 months. Two of them died because of heart failure (6 and 13 months after ICD implantation, respectively). The remote monitoring system reported medical events in 13 (48%) patients. In total, we received 32 event reports (from 1 to 19 per patient, mean 2.6) which were generated due to the detection of ventricular tachycardia (VT) (17 events in 9 patients), ventricular fibrillation (VF) (9 episodes in 6 patients), ineffective defibrillation with the maximal energy (5 reports in 3 patients) and supraventricular tachycardia in the VT detection window (1). Two patients had more than 3 VT/VF episodes during 24 h. There were no reports on technical abnormalities of the ICD system. Interruptions in home monitoring longer than 14 days occurred in 5 (18.5%) patients and lasted 2 to 14 weeks (mean 2.8 ± 7.1). The longest break was caused by the patient’s stay abroad. The remaining interruptions were caused by: journeys (5 episodes), hospitalisations (4), and a temporary stay in a place without sufficient GSM coverage (3). During the follow-up period there were no interruptions in monitoring caused by transmitter or ICD failure. All data received by the home monitoring system were confirmed during the follow-up visits.
Conclusion: Remote monitoring of ICD recipients in Poland does not present technical difficulties and enables early detection of serious events in ICD patients.
Get Citation

Keywords

implantable cardioverter-defibrillators; remote monitoring; telecardiology

About this article
Title

Original article
Technical and practical aspects of remote monitoring of implantable cardioverter-defibrillator patients in Poland – preliminary results

Journal

Kardiologia Polska (Polish Heart Journal)

Issue

Vol 67, No 5 (2009)

Pages

505-511

Published online

2009-05-20

Bibliographic record

Kardiol Pol 2009;67(5):505-511.

Keywords

implantable cardioverter-defibrillators
remote monitoring
telecardiology

Authors

Andrzej Przybylski
Joanna Zakrzewska-Koperska
Aleksander Maciąg
Paweł Derejko
Michał Orczykowski
Łukasz Szumowski
Franciszek Walczak

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., Świętokrzyska 73 street, 80–180 Gdańsk, Poland

tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail: viamedica@viamedica.pl